EasyManuals Logo
Home>Agilent Technologies>Test Equipment>54641A

Agilent Technologies 54641A User Manual

Agilent Technologies 54641A
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #176 background imageLoading...
Page #176 background image
5-6
Making Measurements
To reduce the random noise on a signal
To reduce the random noise on a signal
If the signal you are applying to the oscilloscope is noisy, you can set up the
oscilloscope to reduce the noise on the displayed waveform. First, you stabilize
the displayed waveform by removing the noise from the trigger path. Second,
you reduce the noise on the displayed waveform.
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Remove the noise from the trigger path by turning on either
high-frequency reject or noise reject.
High-frequency reject (HF reject) adds a low-pass filter with the 3-dB point at
50 kHz. You use HF reject to remove high-frequency noise, such as AM or FM
broadcast stations, from the trigger path.
HF Reject
3 dB down point
Pass
Band
0 dB
50 kHz
dc

Table of Contents

Other manuals for Agilent Technologies 54641A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 54641A and is the answer not in the manual?

Agilent Technologies 54641A Specifications

General IconGeneral
BrandAgilent Technologies
Model54641A
CategoryTest Equipment
LanguageEnglish

Related product manuals