3-2
Installation
Unpacking Your Signal Generator
Unpacking Your Signal Generator
1. Unpack the contents of the shipping container.
2. Inspect the shipping container for damage.
If the shipping container is damaged or the cushioning material inside is stressed, keep
them until you have checked the instrument for proper operation.
3. Inspect the signal generator to ensure that it was not damaged during shipment.
If mechanical damage or defects have occurred, notify the carrier as well as Agilent
Technologies. Keep the shipping materials for inspection by the carrier.
WARNING To avoid hazardous electrical shock, do not connect ac power to the
instrument when there are any signs of shipping damage to any
portion of the outer enclosure (cover and panels).
CAUTION Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.