36 U1241A/U1242A User’s and Service Guide
5 Performance Tests and Calibration
Calibration Overview
Closed-case Electronic Calibration
The meter features closed- case electronic calibration. No internal
mechanical adjustments are required. The meter calculates correction
factors based upon the input reference value you set. The new correction
factors are stored in non- volatile memory until the next calibration
adjustment is performed. Non- volatile EEPROM calibration memory does
not change when power has been off.
Calibration Interval
A 1- year interval is adequate for most applications. Accuracy
specifications are warranted only if adjustment is made at regular
calibration intervals. Accuracy specifications are not warranted beyond the
1- year calibration interval. Agilent does not recommend extending
calibration intervals beyond 2 years for any application.
NOTE
Ensure that you have read Test Considerations before calibrating the meter.

Table of Contents

Other manuals for Agilent Technologies U1241A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1241A and is the answer not in the manual?

Agilent Technologies U1241A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1241A
CategoryMultimeter
LanguageEnglish

Related product manuals