Performance Tests and Calibration 5
U1241A/U1242A User’s and Service Guide 37
Adjustment is Recommended
Specifications are only guaranteed within the period stated from the last
adjustment. Whatever calibration interval you select, Agilent recommends
that complete re- adjustment should always be performed at the calibration
interval. This will assure that the U1241A/U1242A will remain within
specification for the next calibration interval. This criteria for
re- adjustment provides the best long–term stability.
Performance data is measured during Performance Verification Tests and
does not guarantee the meter will operate within the test limits unless the
adjustment is performed.
Refer to Calibration Count to verify that all adjustments have been
performed.