Service 4
U2000 Series Operating and Service Guide 65
Disassembly and Reassembly Procedure
Disassembly Procedure
Disassemble the power sensor by performing the following steps:
Tabl e 4- 11 Disassembly Procedure
CAUTION
Disassemble the power sensor only in a static free workstation. Electrostatic discharge renders the
power sensor inoperative.
1 Remove the top label.
2 Loosen three screws by using M2
to remove the housing.
3 Replace the defective sensor
module with a new sensor
module. Please refer to
Table 4- 10.