Programming Ramp Mode
Some users find it comforting to acclimate
to CPAP therapy each session. This can be
accomplished by programming the Ramp
Mode which allows you to start with a low initial
pressure and gradually increase to the prescribed
treatment pressure.
Setting Ramp Time
When the Z2 is in standby mode, press and hold
the Up Arrow Button for approximately 3 seconds
until the “00 MINUTES” value flashes. While it is
flashing, repeatedly press either the Up or Down
Filter Replacement
Fault Conditions
Inspections
Before each use carefully inspect the Z2, CPAP Tube, Filter, and mask for signs of wear, damage
or discoloration. Contact your retail medical supplier or at www.smallcpap.com for replacement
parts.
Cleaning
Regular cleaning should be performed in accordance with the instructions below.
WARNING: This device can expose you to electrocution. Do not
immerse the Z2, Power Adapter, Power Cord or Battery in water. Always
unplug these devices before cleaning (only as instructed in this User Guide)
and make sure they are dry before plugging back in.
CAUTION: Do not use bleach, chlorine, alcohol, or aromatic-based
solutions, moisturizing or antibacterial soaps or scented oils to clean the
Z2. Never attempt to clean the Power Adapter, PowerShell, Battery or
accessories with any of these solutions. These solutions may create a shock
hazard, cause damage and reduce the life of these products.
Daily
1. Remove the CPAP Tube, Q-lite and the mask from the Z2.
2. Wash the CPAP Tube in warm water using mild detergent.
3. Rinse thoroughly, hang, and allow to dry.
4. The Q-lite has no owner servicable parts to replace. The foam inside the Q-lite has
been tested and found to have no significant bacterial growth after 30 days of use in
a dry environment. The use of a water-based humidifier may affect this performance.
Examine the Q-lite for cracks or breaks in the plastic and replace as neccessary.
5. Before next use, reconnect the Q-lite, CPAP Tube and the Tube Adapter to the Z2 and mask.
NOTE: Soaking the CPAP Tube in water for longer than 10 minutes may
shorten the life of the tube.
Every 90 days
The CPAP tube should be replaced every 90 days or earlier if it becomes discolored, cracked or damaged.
CAUTION: Do not use bleach, chlorine, alcohol, or aromatic-based
solutions, moisturizing or antibacterial soaps or scented oils to clean
the CPAP tube. Never attempt to clean the CPAP tube with any of these
solutions. These solutions may cause damage and reduce the life of the
CPAP tube.
Up
Arrow
Button
Down
Arrow
Button
Ramp
Time
Ramp
Pressure
Figure 10
Arrow Buttons to increase or decrease the ramp time in 5 minute increments. (Figure 10) Set the
ramp time to “00 MINUTES” to disable the Ramp Mode.
When the desired ramp time value is reached, press and hold the Start/Stop Button for
approximately 3 seconds to save your ramp time setting (this setting will be saved even when
the Z2 is not powered). The Z2 will return to standby mode. To return to standby mode without
saving the new setting, press and release the Start/Stop Button momentarily instead of holding it
down.
NOTE: Ramp Time must be set to a minimum of 5 minutes prior to
attempting to program Ramp Start Pressure or the change to Ramp Start
Pressure will not take effect and “FAIL 2” will be shown in the General Status
Display field.
Setting Ramp Start Pressure
Press and hold the Down Arrow Button for 3 seconds until the ramp start pressure value flashes.
While flashing, repeatedly press either the Up or Down Arrow Buttons to increase or decrease the
ramp start pressure by 0.5 cm H
2
O increment. (Figure 10) Ramp start pressure must be lower than
prescribed pressure.
When the desired value for ramp start pressure is reached, hold down the Start/Stop Button for
approximately 3 seconds to save your setting. The Z2 will return to standby mode. To return to
standby mode without saving the new setting, press the Start/Stop Button momentarily instead
of holding it down.
User Software
Breas Medical provides User Software for data viewing which may be downloaded from the Breas
Medical website to any Windows computer. A separate user guide for this software is available for
download at the Breas Medical website, www.breas.us.
Care
The Z2 System needs periodic care. It is the users responsibility to be aware of the following items
for care:
• Inspections
• Cleaning