12. Quali
ty Check
Descripti
on
This tes
t verifies assays measured and recorded
on the instru
ment are accurate.
Process
To verify
that your instrument is
calibr
ated
corr
ectly
Step
Action
1
Set APPLI
CATION
to the o
ption applicable to the check sample supplied with the
instru
ment and Method to Automatic Calibration Selection.
2
From th
e
Read
y to Test
screen,
measure the supplied
check sample 1
0 times.
3
Average the 10 assays. See Averaging, page 58.
4
Compar
e the average measured value
for each
element against the Acceptance Limit
maximum
and minimum on the Check Sample Report supplied with your instrument.
5
If all av
erage
d measurem
ents fall within MAX and MIN limits, continue to use the
instru
ment.
If averages fall outside MAX and MIN limits, with respect to standard deviation, check –
Measurement times
Sample surface
Instrument window
If the instrument does not perform as expected, contact Bruker at
support.hmp@bruker.com, +1 (509) 783-9850.
Indi
vidual
reports
The followi
ng figure is an example only.
When p
erforming a quality check, use the
Check
Sample
Report inclu
ded with your instrument. It contains values specific to it as determined during
testing at the factory.
Check Sample Report, example only
Your Complete Source for
Testing Equipment
www.BergEng.com
Berg Engineering & Sales Company, Inc.
1-847-577-3980
Info@BergEng.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker CTX and is the answer not in the manual?

Bruker CTX Specifications

General IconGeneral
BrandBruker
ModelCTX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals