Bruker Daltonik GmbH 2 Introduction
2.3 Tools and Materials Required
Item Description
Key For instrument key-switch and to acti-
vate top cover lift (delivered with the
instrument).
Allen key 5 mm For loosening and removing source
door screws.
HPLC- grade ethanol
(approx. 200 mL)
For cleaning source parts
Safety wash bottle For example, VITLAB, VITsafe,
narrow mouth, 500 ml PELD (Bruker
#(8)201106).
Safety glasses Safety glasses with side protection.
Tissues Free of any impregnation (for exam-
ple, Kimberly- Clark KIMTECH
SCIENCE Laboratory Tissues;
Bruker #(8)321559
Swab (for cleaning
optics only)
Lint- free; not a Q- tip (cotton bud).
For example, Anti- Static swab
TX761D, from basan (Bruker
#(8)270758
Brush (for cleaning
metal parts only)
Bruker # 228209
Nitrile gloves For example, KIMBERLY- CLARK
PURPLE NITRILE Exam Gloves
(Bruker #(8)267458)
Compressed clean gas
(nitrogen)
For ventilating the system and drying
components
ultraflex Series Cleaning Manual Revision 2 Page 17 of 44