4 Cleaning the Source Bruker Daltonik GmbH
Figure 4-4 Cleaning the laser lens
with a swab
Lens and mirror surfaces can also be
cleaned with a swab . Start in the center,
then move outwards using a circular
action.
Figure 4-5 Drying the lens with a
tissue
Remove any excess ethanol using a
tissue. Roll the tissue into a tight cylinder
so that pressure can be applied to the
surface.
4.1.3 Mirror Plate
Figure 4-6 Flushing the ion optic mirrors
using ethanol
Flush the ion optics' mirrors with ethanol using the
wash bottle. See Figure 3-14 for location of the
mirror plate.
Figure 4-7 Cleaning the ion
optic laser mirror
Clean the ion optics laser mirror
with a swab soaked in ethanol.