Bruker Daltonik GmbH 4 Cleaning the Source
Flush the upper mirror (laser mirror) with ethanol to clean the surface (flush upwards into
the metal prism to reach the mirror surface
Note The arrow in Figure Figure 4-1 indicates the tip of the wash bottle spout.
4.1.2 Laser Lens
Swabs are disposable. Always use a fresh swab for each cleaning procedure.
Note Always soak cleaning materials in ethanol. Do not clean parts with dry tissues,
swabs, or brushes.
Figure 4-2 Flushing the upper
surface of the laser lens
Flush the upper surface of the laser lens
above the ion optics with ethanol using
the wash bottle.
Note The arrow indicates the tip of
the wash bottle spout.
Figure 4-3 Flushing the underside
of laser lens
Then, flush the laser lens from
underneath.
Note The arrow indicates the tip of
the wash bottle spout.
ultraflex Series Cleaning Manual Revision 2 Page 31 of 44