28
Setting up
Remove the items packed separately (accessories) from the main carton, and
check them for completeness.
4 fastening screws with washers.
1 Operating instructions.
The device should be set up in a location not exposed to humidity and damp.
The setting up location must be clean, dry and well-ventilated. During opera-
tion, the housing may heat up to approx. 80° C. Therefore maintain a minimum
clear distance of 100 mm and make sure that the ventilation slots are not cov-
ered. Attach the device securely with the four fastening screws supplied.
Important
Make sure that the ventilation slots remain unobstructed! The minimum clear
distance must be 100 mm all around! Insufficient ventilation can cause over-
heating of the device! The device is designed for operation in ambient tem-
peratures of up to 35° C. If the temperature inside the device rises as a result of
insufficient air circulation or of an excessively high surrounding temperature,
the controller will reduce the voltage.
Removing the cover of the terminal area
The terminal area cover must be removed in order to exchange one of the flat
plug-in fuses or in order to connect or disconnect cables. Press the lock, 7 (see
Figure 1) inwards until it unlatches. Lift up the terminal area cover until it is
possible to pull it out forwards. Withdraw the terminal area cover.
The terminal area cover is replaced by reversing these steps. Ensure that the
pegs of the terminal area cover are pushed into the slots in the rear housing
cover, and that the lock at the front has engaged.