4
Important Safeguards
Instrument Safeguards
• Service can only be performed by an authorized CEM service technician. This instrument utilizes high voltages
and microwave radiation.
• DO NOT tamper with the instrument.
• DO NOT install the MARS 6 instrument inside of a fume hood. The exhaust hose must be connected and draw at
least 125 CFM at the point of connection at all times as it is essential for removing harmful gases away from the
instrument. Vapors should be vented into a fume hood by means of the exhaust hose only.
• If any damage to the instrument is noted, do not attempt instrument operation.
• Use only CEM specied consumables and accessories.
• If installed, remove the ESP-1500 Plus cable connection and clean as outlined in this manual.
• Clean the temperature sensors as outlined in this manual.