3. Specification
3.1 Accessories
No DESCRIPTION Q’TY
1 Electrical stimulator device 1 piece
2 Electrode Leads 2 pieces
3 1.5”x 1.5” Adhesive Electrodes 4 pieces
4 9V Alkaline Battery, type 6LR61 1 piece
5 Instruction Manual 1 piece
6 Carrying case 1 piece
3.2 Technical information
Channel Dual, isolated between channels
Power
9.0 V DC-1 Alkaline *6LR61 battery(optional AC Adaptor)
Operating
conditions
5°C to 40°C (41 to 104℉ ℉)with a relative humidity of
30%-75%,atmospheric pressure from 700 to 1060 Hpa
Storage
conditions
-10°C to 50°C (14℉ to 122℉)with a relative humidity of
10%-90%,atmospheric pressure from 700 to 1060 Hpa
Dimensions 4.5×2.55×0.9 inches(L*W*H)
Weight 0.28 lbs(With battery)
Tolerance
There may be a ±5% tolerance of all setting and ±10%
tolerance of output of intensity.
Electrode
Detection
Function
The amplitude level will be reset to 0mA when the
amplitude level is 12mA or greater and an open circuit at
either channel is detected.
Timer
Adjustable, from 1 to 60 minutes or continuous, Adjustable
in 1minutes each step. Treatment time countdown
automatically.