EasyManuals Logo
Home>Dräger>Measuring Instruments>Polytron 5200

Dräger Polytron 5200 User Manual

Dräger Polytron 5200
31 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
10 Dräger Polytron 5200 / Dräger Polytron 53X0
Operation
3.3.3 Field wiring: power and relay, or power, relay and
remote sensor version
Solid Oval (Power) X1
Terminal 1
(top)
Terminal 2 Terminal 3 Termin al 4 Terminal 5
V + V - V -
4 to 20 mA
Signal
Ground
Dashed Oval (Relay) X2
Terminal
1 (left) 2 3 4 5 6 7
Fault Fault Alarm 2 Alarm 2 Alarm 1 Alarm 1 Ground
Solid Rectangle (Remote) X3
Sensor Wire Color
S1 S2 S3 S4 S5
Sensing Head Polytron SE Ex PR M1 DD 6812711
Sensing Head Polytron SE Ex PR M2 DD 6812710
Sensing Head Polytron SE Ex PR M3 DD 6812718
Sensing Head Polytron SE Ex HT M DD 6812720
Black
Wire S2
to
S1
Yellow
Wire S4
to
S5
Brown
Sensing Head Polytron SE Ex LC M1 DD 6812722
Sensing Head Polytron SE Ex LC M2 DD 6812721
Sensing Head Polytron SE Ex LC M3 DD 6812719
Black
Wire S2
to
S1
Yellow
Wire S4
to
S5
Brown
Sensing Head DrägerSensor IR complete Set e 6811165
Sensing Head DrägerSensor IR complete Set e 6811265
Black
Wire S2
to
S1
Yellow
Wire S4
to
S5
Brown
Sensing Head Dräger PIR 3000 complete Set e 6811160
1
Sensing Head Dräger PIR 3000 complete Set e 6811270
1
----- ----- Brown Black Red
1) not yet available
www.norrscope.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 5200 and is the answer not in the manual?

Dräger Polytron 5200 Specifications

General IconGeneral
BrandDräger
ModelPolytron 5200
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals