Operation in Measurement Mode
25
Operating Instructions Vamos, 1. US ed.
Setting Alarm Limits
Setting Alarm Limits
When switching the monitor on, and after Standby mode, the
default alarm limits listed below are active.
lo = Lower alarm limit
hi = Upper alarm limit
– – = Alarm limit deactivated
The adjustable alarm limits are listed in the right-hand column
of the table:
After standby SpO
2 and Pulse alarm monitoring is disabled
until at least 6 pulses with a minimum pulse rate of 30 1/min
have been detected. The "SpO
2 Sensor?" and "SpO2 INOP"
alarms are always monitored.
After standby the gas concentration related alarms are
disabled until the first complete breath has been detected.
Default value Adjustment range
FiCO
2 hi 5 mmHg
0.7 Vol.%, 0.7 kPa
– –, 1 to 20 mmHg
– –, 0.1 to 2.6 Vol.%, kPa
FetCO
2 hi 50 mmHg
6.6 Vol.%, 6.6 kPa
– –, 1 to 75 mmHg
– –, 0.1 to 9.9 Vol.%, kPa
lo – – – –, 0 to 74 mmHg
– –, 0 to 9.8 Vol.%, kPa
FiHal hi 1.5 Vol.% 0.1 to 7.0 Vol.%
FiEnf hi 3.4 Vol.% 0.1 to 7.0 Vol.%
FiIso hi 2.3 Vol.% 0.1 to 7.0 Vol.%
FiSev hi 3.4 Vol.% 0.1 to 9.9 Vol.%
FiDes hi 12.0 Vol.% 0.1 to 21.9 Vol.%
FiHal lo – – – –, 0 to 6.9 Vol.%
FiEnf lo – – – –, 0 to 6.9 Vol.%
FiIso lo – – – –, 0 to 6.9 Vol.%
FiSev lo – – – –, 0 to 9.8 Vol.%
FiDes lo – – – –, 0 to 21.8 Vol.%
SpO
2 lo 92 % – –, 50 to 99 %
Pulse hi 150 – –, 31 to 300
lo 50 – –, 30 to 299