SD5/SD6 Ultrasonic TableTop Doppler User Manual Product Specifications
- 33 -
Circle Life:
≥300 times
(after 300 circles, only 90% of the capacity can be restored.)
Storage Temperature:
-20°C ∼ +60°C (-4 ºF ~ +140 ºF) (1 month) or
-20°C ∼ +40°C (-4 ºF ~ +104 ºF) (3 months)
8.5 Low Output Summary Table
Low Output Summary Table
(for systems with no transducers having global maximum index values exceeding 1.0)
System: Ultrasonic TableTop Doppler
Transducer Model
(MHz)
I
spta.3
(mW/cm
2
)
TI Type TI Value MI
I
sppa.3
(W/cm
2
)
CW 2.0 (SD5) 1.774
TIS 0.0159
0.00628 0.001774
TIB 0.0301
CW 3.0 (SD5) 5.867
TIS 0.0969
0.00813 0.005867
TIB 0.1113
CW 2.0 (SD6) 3.484
TIS 0.0203
0.00785 0.003484
TIB 0.0531
CW 3.0 (SD6) 5.307
TIS 0.1132
0.00763 0.005307
TIB 0.1141