SD5/SD6 Ultrasonic TableTop Doppler User Manual EMC Information
- 37 -
A2.3 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
The SD5/SD6 is intended for use in the electromagnetic environment specified below. The customer or
the user of SD5/SD6 should assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance level
Electromagnetic environment
-guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the SD5/SD6 , including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
Conducted
RF
IEC
61000-4-6
3 Vrms
150 z to 80 MHz
3Vrms
0.1 Vrms
1.995MHz~2.010MHz
2.985MHz~3.015MHz
d=1.2
P
d=35
P
1.995 MHz to 2.010 MHz
d=35
P
2.985 MHz to 3.015 MHz
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d=1.2
P
80 MHz to 800 MHz
d=2.3
P
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.