FLIRMR160USERMANUALDocumentIdentifier:MR160‐en‐US_AD
22
6.3 UpdatingtheMR160firmware
TheMR160firmwarecanbeupdatedinthefieldbytheuser,withouttheneedforsendingtheunit
inforservice.Ifassistanceisrequired,theusercancontactaFLIRtechnicalspecialist(seeSection
8).Firmwareupdatesprovideperformanceenhancementsandnewfeaturesandfunctions.
Toupdatethefirmwarethefollowingisrequired:
Accesstothewebsitewheretheupdatefile
(s)arelocated:
TheMR160tobeupdated
Theupdatefiles
Followthestepsbelow:
1. Visit tocheckforthelatestupdates.
2. Selectthe‘Downloads’tabandthenselect‘InstrumentFirmware’(Testand
Measurement)fromthedropdownmenu.
3. SelectMR160fromthesec
onddropdownmenu.
4. Selectand
downloadtheupdatedfirmware.
5. Turnthemeteron,andconnectittoacomputerviaUSB.
6. Copytheupdatefile(s)intotherootdirectoryoftheMR160drive(e.g.
MR160_v1_962.hex).
7. DisconnectthemeterfromtheUSBport.
8. Holdthepowerbuttonforonesecondtoturnthemeteroff.
9. Pressthepowerbuttontoturnthemeteronagain.Thescreenshouldremaindark;the
unitwillbeeptwicetosignalthattheupdatehasstarted.
10. After~10sitwillbeepagai
nsignalingthattheupdateiscomplete.
11. Pressthepowerbuttontoturnthemeteron;thenewversionshouldnowberunning.
12. Ifthereisanerror,repeattheprocedure.Ifaproblempersists,contactFLIRtechnical
support.