2-20 Fresenius Medical Care 4008 4/09.03 (TM)
3 Heat sink (power supply unit)
4 Heater rod housing (hydraulic unit open)
5 Dialyzer line ports (adapters)
Legend
1 Ground stud for potential equaliza-
tion
2 Upper rear panel (screw)
Test 4008 B, 4008 S
1. Protective earth resistance measuring points
2. Use a meter (e.g. SECUTEST 0701) to check the leakage current.