Fresenius Medical Care 4008 4/09.03 (TM) 8-35
8.9 P.C.B. LP 624 Control board (BP)
8.9.1 Description
General information
This printed circuit board comprises both the control and the power part.
Plug connections on P.C.B. LP 624:
X186, connection to the position sensor
X188, connection to the stepper motor
X189, connection to P.C.B. LP 748
X190, connection to the lid switch
X192, connection to the pressure transducer
X348, connection to the dialysis monitor
Hex switches on P.C.B. LP 624:
0 Arterial blood pump
1 Single needle blood pump
2 – A Without function
B Adjustment of BP stop alarm (15 or 30 sec)
C Without function
D Inquiry of time meter (display x 100 = number of hours)
E Test operation (for the manufacturer only)
F Calibration of pressure transducer
Voltage generation
Both the +24-V and the +12-V voltage supply are made available to the blood pump by the
monitor. From the +24-V voltage, the +5-V supply voltage is generated on the module by the
switching regulator IC 20, in order to minimize the dissipated power.
Stepper motor activation
For noise reduction purposes, the stepper motor is operated in the microstep mode. Resolution
amounts to 60 microsteps per step. The RISC processor alternately transmits an 8-bit dataword
to pins 3 and 5 of the two D-A converters of IC 7. At the output of the converters, two sine-wave
voltages are available, which are phase-shifted by 90 degrees. Together with the current
direction signals, these voltages are supplied to the stepper motor controller IC 2. Together with
the two SM drivers IC 1/IC 22 and the current sensor resistors R58/R59, IC 2, as feedback, forms
a closed control loop, which impresses two sine-wave currents (phase-shifted by 90 degrees) in
the two windings of the SM.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 B and is the answer not in the manual?

Fresenius Medical Care 4008 B Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 B
CategoryMedical Equipment
LanguageEnglish

Related product manuals