1-78 Fresenius Medical Care 4008 4/08.03 (TM)
1.2 Functional description of the modules
1.2.1 Blood pump (arterial)
The blood pump ensures a sufficient blood flow in the extracorporeal blood circuit. It is absolutely
necessary that sterility is maintained and that the blood is prevented from becoming contaminat-
ed.
The blood pump is designed as roller pump provided in an exchangeable plug-in module
integrated in the hemodialysis machine. The blood line is installed between a stator, which, with
its rolling surface bent in a circle, represents a thrust bearing, and a rotor, which is provided with
rollers and pivoted in the stator. The pressure of the rollers causes the development of a narrow
or seal. If the rollers are moving in the direction of delivery, the blood is pushed in this direction.
A microprocessor controls the stepper motor with quartz accuracy, depending on the selected
delivery rate, the set line diameter, and the monitor signals.
The pressure measuring equipment comprises a piezo-resistive pressure transducer. The pres-
sure-proportional voltage is indicated on the monitor on a quasi-analog LED scale.
Functions of the blood pump:
– RAM and CRC test after turning power on,
– control and monitoring of the function by a dual processor system,
– emergency switchoff in case of an alarm: stop recognition (15 or 30 sec),
– setting of the speed to 180 ml/min during priming,
– measurement of the arterial pressure or the single needle pressure (depending on the model
concerned),
– semi-automatic loading and unloading of the line segment.
Error messages:
E.01 Line diameter outside the permissible range
E.02 Undefined hex switch position
E.03 Uncalibrated arterial pressure transducer
E.04 Run-time monitoring error during SN operation
E.05 SN stroke volume outside the permissible range
E.06 SN pressure thresholds outside the range of values of the A-D converter
E.08 Stop alarm
E.09 Error during A-D conversion
E.12 Rotary monitoring error (Hall sensor)
E.13 Monitoring error with regard to current sensing resistors
E.14 Monitoring error with regard to current sensing resistors
E.15 Speed monitoring error

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 B and is the answer not in the manual?

Fresenius Medical Care 4008 B Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 B
CategoryMedical Equipment
LanguageEnglish

Related product manuals