EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>4008 B

Fresenius Medical Care 4008 B User Manual

Fresenius Medical Care 4008 B
595 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #208 background imageLoading...
Page #208 background image
Fresenius Medical Care 4008 4/08.03 (TM) 5-9
5.4 Reading the analog inputs of CPU II
Explanation:
UF Volume display: ADC value
Time Left display: Analog voltage (in 0.1 V), converted to the value at the input of P.C.B. LP 632
䊲
䊲
䊱䊲
䊱䊲
䊱䊲
䊱䊲
䊱䊲
䊱䊲
䊱䊲
䊱䊲
䊱䊲
䊱䊲
䊱䊲
䊲
䊲
CPU2: RD ANALOG INP.
Confirm
key
I: CPU2_BPR_ART
Arterial blood pump
0 to 9.6 V, IC 12/20, ADC 0 to 223
I: CPU2_P_ART
Arterial pressure
0 to 10.6 V, IC 12/19, ADC 0 to 245
I: CPU2_P_VEN
Venous pressure
0 to 11 V, IC 12/18, ADC 0 to 255
I: CPU2_P_DIAL
Dialysate pressure
0 to 10 V, IC 12/17, ADC 0 to 231
I: CPU2_COND_SIGNAL
CD display
0 to 10.8 V, IC 12/16, ADC 0 to 251
I: CPU2_TEMP_DIAL1
Temperature display
0 to 10.8 V, IC 12/15, ADC 0 to 251
I: CPU2_P_DIAL2
Control voltage for higher resolution
Dialysate pressure
0 to 10.9 V, IC 12/14, ADC 0 to 252
I: CPU2_BLL_DIM
Blood leak dimness voltage
5.0 V, IC 12/13, ADC 116
I: CPU2_BLL
Blood leak voltage
5.0 V, IC 12/13, ADC 116
I: CPU2_+10V
Reference voltage, D-A converter/CPU II
10 V; 12/13, ADC 234
I: CPU2_NC6
Not used
0 V, IC 12/13, ADC 0
back to menu ?
Confirm
key
CPU2: RD ANALOG INP.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 B and is the answer not in the manual?

Fresenius Medical Care 4008 B Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 B
CategoryMedical Equipment
LanguageEnglish

Related product manuals