CPU1: WR DIGIT. OUTP
Confirm
key
O: CPU1_V24
Valve 24
IC 10/19
O: CPU1_V24B
Valve 24b
IC 10/13
O: CPU1_V26
Valve 26
IC 10/18
O: CPU1_V130
Valve 130
IC 10/17
O: CPU1_V30
Valve 30
IC 7/16
O: CPU1_V31
Valve 31
IC 12/19
O: CPU1_V32
Valve 32
IC 12/18
O: CPU1_V33
Valve 33
IC 12/17
O: CPU1_V34
Valve 34
IC 12/16
O: CPU1_V35
Valve 35
IC 12/15
O: CPU1_V36
Valve 36
IC 12/14
O: CPU1_V37
Valve 37
IC 12/13
O: CPU1_V38
Valve 38
IC 12/12
5.9 Writing the digital outputs of CPU I
Explanation:
UF Rate display: 0000 = not active
1111 = active (P.C.B. LP 634 level)
(can be changed with: 4008/E/B: ▲▼ UF Rate, 4008 H/S: +/–)
䊱䊲
O: CPU1_V41
Valve 41
IC 7/13
(After a short time interval, the valve
closes automatically, to prevent the
water from overflowing.)