Fresenius Medical Care 4008 4/08.03 (TM) 1-77
ONLINE module errors
Error message Description
ONL Error F00 Online module error
ONL Error F01 Watchdog error
ONL Error F02 Watchdog error
ONL Error F03 Watchdog error
ONL Error F04 Error in the program sequence
ONL Error F05 +24V_WD dropped to less than 17V or was switched off
ONL Error F06 Time-out of the communication watchdog exceeded
ONL Error F07 A transmission from the module was not confirmed by the dialysis
system
ONL Error F08 General valve error
ONL Error F09 T1 test skipped
ONL Error F10 T1 test for ONLINEplus failed to be passed
ONL Error F11 Reference voltage is outside the tolerance
ONL Error F12 CRC error
ONL Error F13 EEPROM error
ONL Error F14 The monitor disabled the ONLINEplus module
ONL Error F16 Valve error ONL1
ONL Error F17 Valve error ONL1
ONL Error F18 Valve error ONL1