3. Using a 8mm wrench, remove the nut and 2 lock washers that secure the ground wire. Cut the wires to the
heater, and remove and discard the female half of the connector.
4. Remove the 2 screws and lock washers (2.5 mm hex key) that secure the reector and remove the heater
assembly.
5. Remove the 2 nuts and lock washers (5.5 mm wrench) from the back of the reector and remove the
heating element and heater guard.
6. Install the new heating element and reinstall the guard in the reector. Thread the wires from the new
heating element through the grommet in the heater housing.
7. Use the 2 mounting screws and lock washers to reinstall the reector.
8. Mount the new female half of the connector in its mounting. Push the pins from the heating element into
the connector. Route the ground wire straight up from the ground stud and secure it with the lock washers
and nut. Push any slack in the heater wire into the housing.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved. 6600-0343-000 104 115
Chapter 5: Repair Procedures

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed and is the answer not in the manual?

GE Giraffe OmniBed Specifications

General IconGeneral
BrandGE
ModelGiraffe OmniBed
CategoryMedical Equipment
LanguageEnglish

Related product manuals