GE
D
IRECTION 5273008-100, REVISION 9 LOGIQ C SERIES BASIC SERVICE MANUAL
8-2 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Section 8-2 DISASSEMBLY/RE-ASSEMBLY
8-2-1 Warning and Caution
8-2-2 Air filter and Knob Screw (FRU No. 5030)
8-2-2-1 Tools
Common phillips screwdrivers
8-2-2-2 Preparations
Shut Down the System and disconnect the power cord.
8-2-2-3 Removal Procedure
1.) Unscrew the knob screw which fix air filter on the edge of power box.
2.) Pull out the air filter.
8-2-2-4 Mounting procedure
Install the new parts in the reverse order of removal.
WARNINGWARNING
ONLY QUALIFIED SERVICE PERSONNEL SHOULD REMOVE ANY COVERS OR
PANELS. ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE. BECOME
THOROUGHLY FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT
LEVELS TO AVOID ACCIDENTAL CONTACT
CAUTION
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
WARNINGWARNING
DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU UNIT LEVEL AT ANY
CIRCUMSTANCES.
Figure 8-1 Air filter and Knob Screw Disassembly
1)
2)