EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ C Series

GE LOGIQ C Series User Manual

GE LOGIQ C Series
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
GE
D
IRECTION 5273008-100, REVISION 9 LOGIQ C SERIES BASIC SERVICE MANUAL
4 - 26 Section 4-3 - General Procedure
4-3-14 Backup and Restore Database, Preset Configurations and Images
4-3-14-1 Formatting Media
NOTE: The system may crash if use a DVD for Archive/Backup/EZBackup/Restore, so always use a
CD-R or USB HDD for Archive/Backup/EZBackup/Restore. DVD only can be used for upgrade.
1.) To format the backup media, select the UTILITY
button on the Keyboard.
2.) Select CONNECTIVITY
, then REMOVABLE MEDIA. Properly label and Insert the backup
media.
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure 4-16. It is best to use all capital letters with no
spaces or punctuation marks. Press Format
.
5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-17. When the formatting
has been completed, press OK
to continue.
6.) If desired, verify that the format was successful by returning to Utility
->Connectivity-
>Removable Media
and selecting VERIFY as shown in Figure 4-16.
Figure 4-16 Format and Verify Media
Figure 4-17 Format Successful Pop-up Menu

Table of Contents