GE
D
IRECTION 5273008-100, REVISION 9 LOGIQ C SERIES BASIC SERVICE MANUAL
Chapter 1 Introduction 1-7
1-3-3 Mechanical Safety (cont’d)
NOTE: Special care should be taken when transporting the unit in a vehicle:
Secure the unit in an upright position.
Lock the wheels (brake)
1-3-4 Electrical Safety
To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. The power outlet used for this equipment should not be shared with
other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
1-3-5 Label Locations
See LC2/C2 PRO, LC3/LC3 PRO, LC5/C5PRO Basic User Manual (5265060-100,5268001-
100,5268596-100) section “Warning Label Location” for more Information.
WARNINGWARNING
DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU UNIT LEVEL AT ANY
CIRCUMSTANCES.