GE LOGIQ F SERIES
DIRECTION 5446617-100, REVISION 10 BASIC SERVICE MANUAL
Section 10-4 - Tools Required 10 - 7
10-4-3 Input Power
10-4-3-1 Mains Cable Inspection
10-4-4 Cleaning
10-4-4-1 General Cleaning
NOTE: For your convenience or of the air filter is too dirty, replacement filters are available. refer to
Chapter 9 for the air filter replacement part number.
10-4-5 Physical Inspection
Table 10-6 Mains Cable Inspection
Step Item Description
1 Unplug Cord
Disconnect the mains cable from the wall and system.
2Inspect
Inspect it and its connectors for damage of any kind.
3Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no
strands may cause a short circuit.
4 Verify
Inlet connector retainer is functional.
Table 10-7 General Cleaning
Step Item Description
1 Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
not to get the cloth too wet so that moisture does not enter the console.
2 Probe Holder
Clean probe holders (they may need to be soaked to remove excess gel).
Table 10-8 Physical Checks
Step Item Description
1 Labeling
Verify that all system labeling is present and in readable condition. refer to the LOGIQ C3/C5
Premium User Manual for details.
2 Scratches & Dents
Inspect the console for dents, scratches or cracks.
3 Wheels & Brakes
Check all wheels and casters for wear and verify operation of foot brake, to stop the unit from moving,
and release mechanism. Check all caster locks and caster swivel locks for proper operation.
4
Cables &
Connectors
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
special attention to footswitch assembly and probe strain or bend reliefs.
5
Shielding &
Covers
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems while scanning.
6 External I/O
Check all connectors for damage and verify that the labeling is good.
7 Op Panel Lights
Check for proper operation of all operator panel and TGC lights.
8 Monitor Light
Check for proper operation of any monitor lights if available.
9
External
Microphone
Check for proper operation of any external microphones by recording an audio test.