EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ F5

GE LOGIQ F5 User Manual

GE LOGIQ F5
167 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
GE LOGIQ F SERIES
DIRECTION 5446617-100, REVISION 10 BASIC SERVICE MANUAL
Section 10-4 - Tools Required 10 - 9
10-4-7 Grounding Continuity
Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
The ground wire resistance should be less than 0.2 ohms. Reference the procedure in the IEC 601-1.1.
10-4-7-1 Meter Procedure
Follow these steps to test the ground wire resistance.
1.) Turn the LOGIQ C3/C5 Premium unit OFF.
2.) Plug the unit into the meter, and the meter into the tested AC wall outlet.
3.) Plug the black chassis cable into the meter’s “CHASSIS” connector and attach the black chassis
cable clamp to an exposed metal part of the LOGIQ C3/C5 Premium unit.
4.) Set the meter’s “FUNCTION” switch to the RESISTANCE position.
5.) Set the meter’s “POLARITY” switch to the OFF (center) position.
6.) Measure and record the ground wire resistance.
CAUTION
Electric Shock Hazard. The patient must not be contacted to the equipment during this test.
Figure 10-2 Ground Continuity Test

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ F5 and is the answer not in the manual?

GE LOGIQ F5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ F5
CategoryMedical Equipment
LanguageEnglish

Related product manuals