GE LOGIQ F SERIES
DIRECTION 5446617-100, REVISION 10 BASIC SERVICE MANUAL
Section 2-3 - Facility Needs 2 - 9
2-3-4 Networking Pre-installation Requirements
2-3-4-1 Stand Alone Scanner (without Network Connection)
None.
2-3-4-2 Scanner Connected to Hospital’s Network
Supported networks:
Wire LAN
2-3-4-3 Purpose of DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at
a lower cost.
2-3-4-4 DICOM Option Pre-installation Requirements
To configure the LOGIQ F Series to work with other network connections, the site’s network
administrator must provide some necessary information.
Information must include:
A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ F Series.
The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQ F Series for DICOM APPLICATION INFORMATION. A field for the make (manufacturer)
and the revision of the device, is also included. This information may be useful for solving errors.