Important Servo-i Ventilator System
Revision 02 Service Manual 1 - 7
Articles of consumption
1. Bacteria filter
2. Filters for the gas modules
3. Filter for the inspiration pressure transducer
4. Filter for the O
2
cell
5. Nozzle units for the gas modules
6. Battery modules
7. Lithium batteries
8. Expiratory cassette
9. Expiratory cassette membrane
10. O
2
cell
11. Backlight lamps.
Item 1: Consumption approximately 250 pcs/year.
Items 2 – 5: Changed approx. every 5.000 hours.
Items 6 – 7: Changed approx. every 15.000 hours.
Items 8 – 11: Changed when needed.
Power consumption
The power consumption depends on the operating
mode and whether the internal batteries are being
fast or trickle charged.
Mode Fast charging Trickle charging
In operation 70 W 38 W
Standby 65 W 33 W
Off 35 W 6 W
Noise level
Less than 50 dBA.
Packing materials
The amounts of packing materials will vary
depending on customer adaptation.
Materials for packing:
• Loading pallet. Fulfils the USA requirements
7 CFR 319.40 May 25’th 1995.
• Corrugated cardboard
• Stretch film of Polyethylene, PE.
• Shock-absorbing material of expanded
polyethylene, EPE, or expanded polypropylene,
EPP.
• Clamps of Polyethylene, PE.