Chapter 5, Installation and Configuration 5-15
PCS2 Machine Transfer Verification Form
Customer Name: City/State:
PCS-2 Serial #
VALVES Functional Yes No
WEIGHER Zero Reading
0 gm ± 1
1000 gm ± 5
PRESSURE DPM Offset (1984 - 2112)
300 mmHg ± 8
CUFF 50 mmHg ± 5
100 mmHg ± 10
DISPLAY Functional Yes No
KEYBOARD Functional Yes No
CENTRIFUGE SPEED Set Speed Actual Speed
7000 RPM ± 100
Cover Lock Functional Yes No
Bowl Vibration Functional Yes No
Spill Detect Functional Yes No
Donor Lights Functional Yes No
TRANSDUCERS OPTICS Aim (Yes if Adjusted) Yes No
Open (1-82 or 0 V ± 0.10)
Grey Card
(2295- 3278 or 2.8 - 4.0 V)
LINE SENSOR Open (2866 - 3276)
Black (0-103)
AIR DETECTOR ACAD Functional Yes No
BLAD Functional Yes No
DLAD-1 Functional Yes No
DLAD-2 Functional Yes No
FANS Functional Yes No
Biohazardous Waste Bag Attached Yes No
Program Revision
Cuff
Weigher
Circle if adjusted DPM
Centrifuge
Optics Voltage
Line Sensor
Performed by: Print Name:
Date: