To examine software execution when a timing
violation occurs
The timing analyzer triggers when the timing violation occurs, and when it
triggers, it also sets its “arm” level to true. When the state analyzer receives
the arm signal, it triggers immediately on the present state.
1
Select the timing analyzer Trigger menu.
2 Define the Edge1 term to represent the control line where the timing
violation occurs.
3 Under Timing Sequence Levels, enter the following sequence
specification:
TRIGGER on “glitch/edge1” 1 time
4
Select the state analyzer Trigger menu and accept the default
(anystate) definition for term a.
5 Under State Sequence Levels, enter the following sequence
specification:
While storing “anystate” TRIGGER on “arm a” 1 time
Store “anystate”
Triggering
To examine software execution when a timing violation occurs
1–23