EasyManuals Logo
Home>HP>Measuring Instruments>16500C

HP 16500C User Manual

HP 16500C
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
Skew Adjustment
You can modify the skew or timing deviation between modules within
the intermodule measurement. This allows you to compensate for any
known delay of the system under test, or to compare two signals by
first removing any displayed skew between the signal channels.
Skew adjustments can correct module delays to within 2 ns of other
modules. Note that the module or channel that is used as the trigger is
normally the reference channel, and adjustments are normally made
to deskew the nonreference channels.
2–26

Table of Contents

Other manuals for HP 16500C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 16500C and is the answer not in the manual?

HP 16500C Specifications

General IconGeneral
BrandHP
Model16500C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals