Technical Specification
Parameter IT6018C-300-225
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Readback Stability-30min Voltage
≤0.02%+0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Readback stability-8h Voltage
≤0.02%+0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Efficiency
~92%
Remote Sense Compensa-
tion Voltage
≤5V
Command Response Time
2mS
Power Factor 0.99
Storage Temperature
-10℃~70℃
Protective Function
OVP, OCP, OPP, UVP, UCP, OTP, Vsense protection
Standard Interface
Standard: USB, CAN, LAN, VCP; optional: GPIB, analog card
(includes RS232), fiber optic socket
Isolation ( output to ground)
1000V
Working Temperature
0~50℃
Dimension ( mm) 483mm(W)*801.61mm(D)*151.3mm(H)
Weight( net)
40KG
Note:
1. Models without emergency stop module (i.e. AC contactor not included), other AC input volt-
age ranges are required for customization. Models without emergency stop module and as-
sembled to the cabinet, with AC input voltage in the range of 198~264V, need to be customized
while derating by 50%. Models with emergency stop module (including AC contactor) only sup-
port 380VAC±10% for standard AC input voltage, 480VAC±10% voltage is required for
customization.
6.1.3 500V
6.1.3.1 IT6006C-500-40
Parameter IT6006C-500-40
Rated value
Output Voltage
0~500V
( 0 ℃-50 ℃) Output Current
-40~40A
Output Power
-6~6kW
Copyright © Itech Electronic Co., Ltd.
210