EasyManuals Logo
Home>ITech>Test Equipment>IT8811

ITech IT8811 User Manual

ITech IT8811
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Function and Features
Copyright © ITECH Electronic Co., Ltd. 16
3.2.4 Constant power operation mode (CW)
Under CW mode, the electronic load will consume a constant power, as shown
below. If input voltage rises, the input current decreases and power P (= V * I)
will maintain at setting power. As shown in Fig. 3-4.
Fig. 3-4 Voltage-Current Relation Schema under CW Mode
Under CW mode, the electronic load provides three ways to modify constant
power.
Rotate adjusting knob to set constant power value.
Use numeric keys to input power value and press [Enter] key to confirm set
constant power value.
Use
moving cursor and press to adjust values at
corresponding positions.
Operation steps
1. Press [CW] key and [Shift] + [CV] to enter parameter setting screen.
Constant Power
Range=400.00W
2. Set maximum working power value and press [Enter] for confirmation.
Constant Power
Range =300.00W
3. Set maximum voltage value and press [Enter] key.
Constant Power
High=130.00V
4. Set minimum voltage value and press [Enter] key.
Constant Power
Low=0.000V
5. Complete parameter setting.
10.0000V 0.0000A
0.00W CW=1.00W
NOTE
If the above method is for editing auto test step (as mentioned below), constant
power range can also be set.
Input
voltage
Load current
V
I
Set power
V
2
V3
I2
I3
CW mode

Table of Contents

Other manuals for ITech IT8811

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8811 and is the answer not in the manual?

ITech IT8811 Specifications

General IconGeneral
BrandITech
ModelIT8811
CategoryTest Equipment
LanguageEnglish

Related product manuals