Function and Features
Copyright © ITECH Electronic Co., Ltd. 25
If the set OCP voltage value is higher than the power voltage value, the OCP will fail
to operate and the board will display as follows:
9.9990V 0.0009A
0.01W 0.100A FAULT STOP
3.11 OPP test function
The IT8800 series electronic load is provided with overpower protection test
function (OPP). Under OPP test mode, when input voltage reached Von value,
delay for a while for the electronic load to latch. Ascend value by step value at
regular interval. At the same time, check the load input voltage and judge
whether it is higher than standard voltage value. If higher, it indicates that OPP
does not occur. Repeat power stepping operation till the load operates to the
cutoff power; if lower, it indicates that OPP does occur. Check whether the
existing power value is within target scope again. If yes, PASS the test.
Press [Shift] + [CW] keys to enter OPP test function setting screen:
OPP
TEST
Run
OPP TEST
Operate OPP test documents
Recall
OPP TEST
OPP TEST
Recall OPP File=1
Recall OPP test documents (1-5)
Edit
OPP TEST
1:Voltage on level=0.000V
Set Von voltage value
2:Voltage on Delay=0.00S
Set Von voltage delay time
3:Current Range=0.000A
Set current range
4:Start Power=0.000W
Set initial power value
5:Step Power=0.000W
Set step power value
6:Step Delay=0.00S
Set step delay time
7:End Power=0.000W
Set cutoff power value
8:OCP Voltage=0.000V
Set OCP value
9:Max Trip Power =0.000W Set overpower range
(maximum value)
10:Min Trip Power =0.000W
Set overpower range
(minimum value)
Save OPP File=1
1-5
Save OPP test documents
Operating steps:
1. Press [Shift] +
key to start OPP test. If within range, PASS the test
and the board will display as follows:
9.996V 0.0007A
0.01W 49.10W PASS STOP
If not, there is FAULT and the board will display as follows:
9.996V 0.0007A
0.01W 48.6W FAULT STOP
2. End test. Press [ESC] to return to setting screen. Press [ESC] again to exit.
NOTE

Table of Contents

Other manuals for ITech IT8811

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8811 and is the answer not in the manual?

ITech IT8811 Specifications

General IconGeneral
BrandITech
ModelIT8811
CategoryTest Equipment
LanguageEnglish

Related product manuals