The wire must be heavy enough not to overheat while
carrying the short-circuit output current of the unit.
Please meet the wiring requirements described above.
Turn on the instrument
Turn on the instrument by pressing the front-panel POWER
switch to the on position. The instrument powers up.
Connections for testing
Use the wire ratings listed in the following table when making
connections to the instrument.
Usage Specications
Front panel binding posts AWG 20 to AWG 12
Rear panel output terminals AWG 20 to AWG 12
Two-wire connections
Two-wire connections are used for basic operation when
maximum accuracy is not required. Keep the wire as short
as possible to reduce lead inductance and noise pickup. The
following illustration shows two-wire (local sensing) DUT
connections to the front panel.