SECTION
2
Calibration
13.
As
performed
in steps
8
and
9,
measure
the
voltage
across
the
lOOk2
resistor
and
caiculate
the
current.
VlookO
=
VlookO
I9A
=
R100kQ
v
22.
Record
the
voltage
measurement
from
the Model
193A
and
caiculate
the
current.
Vko
Vlk2
I9OtA
=
Rlk2
mV
I9RA
IS
the
characterized
current
for
the
2OjiA
range
of
the
Model
263.
lt
is
used
to
calibrate
the
lOjiA
range
of
the
Model
236/237.
Using
a
pendil,
record
the
cal
culated
value
of
I9LA
in
Entry
2
of
Table
2-3.
14.
Place
the Model
263
in
standby.
15.
In
the
test
fixture,
remove
the
lOOk2
resistor
and
in-
stall
the
lk2
resistor
(R-263-lk).
16.
On the
Model
193A,
select
the
Ohms
function,
2k2
range
and
disable
ZERO.
17.
Repeat step
3
to
zero
the Model
193A.
18.
Measure
and
record
the
value
of
the
lk2
resistor.
R
1
=
19.
Set
the
Model
193A
to
DCV
and
the
200mV
range,
and
disable
ZERO.
20.
To
cancel
any
offset,
zero
the
Model
193A
by
short
ing
its
input
(VOLTS
HI
to
LO)
and
enabling
ZERO.
21.
Remove
the
short
from
the
input
of
the
Model
193A
and
set
the
Model
263
to
source
90.000jaA
on
the
200jiA
range
and
enable
OPERATE.
I9OA
15
the characterized
current
for
the
2OOiA
range
of
the
Model
263.
lt
is
used
to
calibrate
the
1OOtA
range
of
the
Model
236/237.
Using
a
pendil,
record
the
calculated
value
of
I9OA
in
Entry
3
of
Table
2-3.
23.
Place
the
Model
263
in
standby.
24.
Set
the
Model
193A
to
the
2V
range.
25.
Set
the
Model
263
to
source
O,90000mA
on
the
2mA
range,
and
enable
OPERATE.
26.
Record
the
voltage
measurement
from
the Model
193A
and
caiculate
the
current.
Vike
VlkO
IO9mA
RlkO
v
IO9mA
]S
the characterized
current
for
the
2mA
range
of
the
Model
263.
lt
is
used
to
calibrate
the
bnA
range
of
the
Model
236/237,
Using
a
pencil,
record
the
calculated
value
of
IO9mA
in
Entry
4
of
Table
2-3.
27.
Place
the
Model
263
in
standby.
28.
In
the
test
fixture,
remove
the
1ki
resistor
and
install
the
1OO2
resistor
(R-249-100),
Table
2=3,
Model
263
ISource
Characterization
263
I-Source
Programmed
Characterized
ISource
Entry
Range
Value
ISource
Value
Value
Limits*
1
2jiA
O,9OOOO.tA
IO
9
O,9iA
±45OpA
2
2OjiA
9.000jiA
I
9
A
9iA
±4.3nA
3
200jiA
90,000jaA
I
9
O
9OjiA
±45nA
4
2mA
O,9000mA
Iou
O,9mA
±43OnA
5
2OmA
9.0000mA
I
9
mA
The
characterized
Isource
values
should
be
within
the
listed
limits,
These
limits
are
based
on
the
90-day
specifications
of
the
Model
263
and
193A.
A
characterized
value
that
does
not
fall
within
these
limits
indicates
a
calculation
error,
or
that
the
Model
263
and/or
Model
193A is
out
of
specification.
25

Table of Contents

Other manuals for Keithley 236

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 236 and is the answer not in the manual?

Keithley 236 Specifications

General IconGeneral
BrandKeithley
Model236
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals