SECTION
2
Calibration
Table
24.
Calibration
Error
Messages
Message
Expianation
IN
CAL
Tried
to
set
the
instrument
to
a
state
that
is
illegal
while
in
the
calibration
mode.
Legal
com
mands
include;
C
(calibration),
N
(operate/standby),
0
(output
sense)
and
U
(status)
From
the front
panel, only
operate/standby
and
output
sense
can
be
set.
All
illegal
commands and
operations
are
ignored.
IN
STBY
The
Model
236/237
is
in
standby.
The
calibration
command
sent
over
the
bus
will
not
be
exe
cuted
(see
Note).
CAL
COMPLI
Compliance
has
occurred
during
calibration. Check
test
setup
connections
and
instrument
settings.
CAL
MODE
ERR
Common
mode
calibration
error.
Output
of
Model
236/237
is
not
shorted
or
unit
is
defective.
The
cal
constant
from the
previous
calibration
will
be
used.
CAL VALUE
ERR
Invalid
calibration
value
sent
to
the Model
236/237.
This
calibration
step
is
ignored
(not
at
tempted). Calibration constant
from
previous
calibration
will be
used.
CAL
ADC
ZERO
Measure
offset
calibration
error.
Calibration
is
attempted
using
a
value
that
is
not
within
ac
ceptable
limits. The
transmitted
value
is
ignored
and
a
default
cal
constant
value
is
used
in-
stead.
Possible
problems
include;
incorrect
cal
constant
value, defective
calibration
equip
ment
(or
connections),
or defective
Model
236/237.
CAL
ADC
GAIN
Measure
gain calibration
error.
Calibration
is
attempted
using
a
value
that
is
not
within
ac
ceptable
limits. The
transmitted
value
is
ignored
and
a
default
cal
constant value
is
used
in-
stead.
Possible
problems
include;
incorrect
cal
constant
value,
defective
calibration
equip
ment
(or
connections),
or defective
Model
236/237.
CAL
SRC
ZERO
Source
offset
calibration
error.
Calibration
is
attempted
using
a
value
that
is
not
within
ac
ceptable
limits.
The
transmitted
value
is
ignored
and
a
default
cal
constant
value
is
used
in-
stead.
Possible
problems
include;
incorrect
cal
constant
value, defective
calibration
equip
ment
(or
connections),
or
defective
Model
236/237.
CAL
SRC
GAIN
Source
gain
calibration
error.
Calibration
is
attempted
using
a
value
that
is
not
within
accept
ahle
limits.
The
transmitted value
is
ignored
and
a
default
cal
constant
value
is
used
instead.
Possible
problems
include;
incorrect
cal
constant
value,
defective
calibration
equipment
(or
_________________
connections),
or
defective
Model
236/237.
Note:
During
the
execution
ofa
calibration
step
in
which
the Model
236/237
takes
a
measurement
conversion,
it
is
possible
the
instrument
can
transition
from
OPERATE
to
STANDBY.
This
can
occur
because
it
was
specifically
programmed
to
do
so,
or
because
the
test fixture safety
interlock
was opened.
In
this
event,
the
OPERATE
to
STANDBY
transition
is
detected
and
the
measurement
conversion
used
to
compute
the
calibration
constant
is
ignored.
The
front
panel message
“IN
STBY“
is
displayed
when
this
occurs.
244

Table of Contents