SECTION
4
Principles
of
Operation
The
voltage
drop
across
the
bad
is
20V
(5niA
x
4k2
=
20V).
This
voltage
is
scaled
down
to
2V
and
invertei
This
2V
is
fed
back
(via
V
FEEDBACK)
to
U15A
and
U5B.
The
non-4nverting
(+)
input
of
U15B
will
go
posi
tive
keeping
CR12
reverse-biased.
As
a
result,
the
+V
CLAMP
remains
off.
Current
from
±10V
REF
will
flow
into
the
+1
CLAMP,
which
is
turned
on
(remember,
the
other
clamps are
off).
As
a
result, the feedback
configuration
and the
1
DAC
output
will
apply
a
voltage
to
the
error
amplifier
(U14)
that
will
effect
a
constant
+5mA
output
current.
Main
To
Output
Stage
Figure
42.
Simplified
Schematic
of
VII
Control
and
Readback
43

Table of Contents

Other manuals for Keithley 236

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 236 and is the answer not in the manual?

Keithley 236 Specifications

General IconGeneral
BrandKeithley
Model236
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals