Model 6514 Quick Results Guide 13
Remote command programming
The SCPI commands for speed, digits, filter, and rel are listed in Table 8. The commands
for ranging are listed in Table 4.
Table 8
SCPI commands: rate, digits, filter, and rel
Command Description
Rate commands:
[:SENSe[1]]:VOLTage:NPLCycles <n> Set voltage rate in PLCs: 0.01 to 10.
[:SENSe[1]]:CURRent:NPLCycles <n> Set current rate in PLCs: 0.01 to 10.
[:SENSe[1]]:RESistance:NPLCycles <n> Set resistance rate in PLCs: 0.01 to 10.
[:SENSe[1]]:CHARge:NPLCycles <n> Set charge rate in PLCs: 0.01 to 10.
Digits command:
:DISPlay:DIGits <n> Set display resolution: 4 to 7.
Median filter commands:
[:SENSe[1]]:MEDian:RANK <n> Set median filter rank: 1 to 5.
[:SENSe[1]]:MEDian[:STATe] <b> Enable or disable median filter.
Digital filter commands:
[:SENSe[1]]:AVERage:TCONtrol <name> Select filter control: MOVing or REPeat.
[:SENSe[1]]:AVERage:COUNt <n> Set filter count: 1 to 100.
[:SENSe[1]]:AVERage[:STATe] <b> Enable or disable digital filter.
Rel command:*
:CALCulate2:NULL:STATe <b> Enable/disable rel.
* Does not include commands to set and acquire rel values. See Section 7 of Model 6514 Instruction Manual.

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals