CDE/CME/CDEE-BA-e-1014
26
9 Data output RS 232 C
9.1 Technical data
• 8-bit ASCII Code
• 1 start bit, 8 data bits, 1 stop bit, no parity bit
• Baud rate selectable at 1200, 2400, 4800, 9600 and 19200 Baud
• Miniature plug-in necessary (9 pole D-Sub)
• For operation with interface faultless operation is only ensured with the correct
KERN – interface cable (max. 2m)
9.2 Pin allocation of the balance output socket (front view)
5 1
9 6
Pin 2: Transmit data
Pin 3: Receive data
Pin 5: Signal ground
9.3 Interface parameter
General Information
The previous condition for the data transfer between balance and a peripherical
device (e.g. printer, PC) is that the appliances are set to the same interface
parameters (e.g. baud rate, transfer mode).
9.3.1 Data transfer mode
Ö With the weighing scale switched off, press ON/OFF and TARE-key
simultaneously until the first function “Unit“ appears.
Ö Repeatedly press the 10x-key until “PR“ appears.
Ö Confirm by pressing the PRINT-key. Current setting appears.
Ö Now you can use the 10x-key to select among the different settings
(see chpt. 8.3)
Ö Changed setting will be imported by pressing the PRINT key. The balance
returns automatically into weighing mode.