7.1 Procedure:
In weighing mode, press repeatedly until [tCH] appears.
Press and [Pin] will appear.
Press , and one after the other and [P1 SPd] will
appear
Press , „P2 CAL“ will be displayed
Operate adjustment switch; for position see chap.16
Press and [dESC] will appear
Press repeatedly until „CAL“ will be displayed.
To confirm, press and [UnloAd] will appear
Ensure that there are no objects on the weighing pan.
Wait until the stability display „STABLE“ is displayed and then
confirm by pressing .
(example)
The size of the currently set adjustment weight will be
displayed.
To change, select the digit to be altered by , and the
numerical value by .
Confirm by pressing and [LoAd] will be displayed.