9
The internal light comes on when the door opens and turns off when it is shut. It can be
changed so the light is on or off all the time. To change the setting, access the menu in the
usual way by holding both the arrow icons/buttons and scroll through the menu until you get
to the Lt screen or hold the light icon/button which will take you straight to the Aut screen.
Touch the tick icon/button as before to access this function and the screen will show Aut. This is the
default setting described above. To change the setting to the light being on all the time touch an
arrow icon/button until the screen displays on. Once selected, hold the tick icon/button until the
display reverts to Lt. To change the light operation to not coming on at all, select OFF instead and
again hold the tick icon/button until the screen reverts to Lt. You can either touch the mute icon/
button or just leave the controller alone and the screen will return to the normal temperature display.
Internal Light
Battery Alarm
The battery alarm is enabled by default. If you put 4AA batteries in your refrigerator the
alarm will sound if the batteries are faulty, removed or there is a mains power failure. Labcold
strongly recommends that you fit batteries to your IntelliCold® pharmacy refrigerator for
maximum cold chain safety, however, should you decide not to, you will need to disable the
alarm please see Turning the Battery Alarm O opposite
The alarm shows on the normal display as a battery symbol which alternates with a screen
showing n-b. The alarm will sound when n-b is displayed. This alarm can be muted by
touching the mute icon/button while the n-b screen is displayed. The visual alarm will remain
and the screen n-b will still alternate with the temperature display. Unless you install batteries
the alarm will sound again after the time out interval is exceeded. (see page 8).
Battery Alarm