Edition 3 April 2012 Mars/Mars Pro Service Manual
Page 15/59
The electric frame is installed in the same way with the exception that the side panels house the electric motors for raising /
lowering the cabinet.
On the rear of the frame is positioned the motor connection box from which the power cable is taken. The power cable
must be connected to a suitable mains voltage power source.
3.3 Media valves (Gas / Vacuum )
The Mars cabinet is designed to accept the fitting of media connections through the
side windows. Both side windows are, as standard, provided with one hole of Ø
23mm for the fitting of such connections. If more media connections are required
then the window can be replaced with one having 3 holes (see Fig 2 item 9).
The media valves are colour coded in accordance with international standards so it is
important to install the correct valve for the purpose required. If flammable gas is required
in the cabinet then the gas line must also be fitted with an electric solenoid valve, which
will cut off the gas supply in the event of a power failure. (Part No 9,000,020,030). This
will be fitted to the inlet pipe on the outside of the cabinet and wired back to
the main control board as detailed in part 3.3.1.