40
6 Correcting Faults
6.1.2 Warnings
Warning no. Warning text
001 (HP)
101 (ÜP)
O
2
offset voltage in the air too large or too small
/1 - 29999 Trigger value of the O
2
offset voltage in the air (x0.1 mV)
003 (HP)
103 (ÜP)
O
2
temperature too large or too small, was not assumed
invalid calibration (p. ex. with calibration gas >14,9 Vol % O
2
). The new
O
2
temperature is not accepted. The O
2
measurement remains/will be
improper.
/1 - 29999 Trigger value of the O
2
temperature (x0.1 K)
010 (HP)
110 (ÜP)
For LT3-F: Internal resistance of ceramics of the O
2
electrode
over 45 Ω
For LT3: Internal resistance of ceramics of the O
2
electrode
over 80 Ω
/0 - 65535 Trigger value of current internal resistance (x0.1 ),
tolerance time of 3 seconds.
016 (HP)
116 (ÜP)
Internal warning
/1 Serial number, customer ID or service password reset.
Contact the manufacturer.
Display if customer ID in LT3 and UI do not correspond.

Table of Contents

Other manuals for Lamtec KS1D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lamtec KS1D and is the answer not in the manual?

Lamtec KS1D Specifications

General IconGeneral
BrandLamtec
ModelKS1D
CategoryTransmitter
LanguageEnglish

Related product manuals