59
DATE
22.12.2003
COMPILER TECO/ATI
REG. CODE
1-5302-467
MODEL N°
50707
DATE OF ISSUE
06-95
REVISION 05
ENDORSED
X
121
120
FUEL SYSTEM
Injector calibration
Connect the injector to a hand pump and adjust if necessary,
modifying the thickness above the spring.
When the spring is replaced, calibration must be carried out at a
pressure higher than 10 bars to counterbalance adjustments while
running.
Spraying and opening pressure
With the pressure gauge closed, press hard on the hand pump at
least 10 times.
Open the pressure gauge and press down on the pump once every
second, while keeping a check on the spraying process and
pressure.
The opening pressure must lie between the two values given in the
table. Jets must be uniform and well distributed.
Nozzles
The set-up between the needle and the guide must leave the
needle free to fall and merely as a result of its own weight, when
lifted 7mm from its seat and rotated in different directions, with the
nozzle kept at a 45° angle.
Rotation of the needle must be completely smooth and uninhibited
by obstacles and malformations.
Moreover, on being squeezed against its seat, it must fall freely,
when the nozzle is inverted.
The test must be carried out after rinsing both the needle and
nozzle with trichloroethane and wetting with filtered SHELL
CALIBRATION FLUID “C” oil.
522DL51513DL51
513DL51
APE
053DL51APE053DL51004DL51
004DL51
APE
044DL51
A
5,45,45,45,45,45,45,45,4
B
selohfo.maiddna°N(
71,0x402,0x4141,0x522,0x4051,0x5951,0x5051,0x5661,0x5
htgneleloh5,06,08,06,08,05,08,05,0
αα
α
αα
°061°061°551°061°551°061°061°061
thgiehnip51,0÷01,051,0÷01,0522,0÷571,0571,0÷521,0522,0÷571,0524,0÷573,0524,0÷573,0524,0÷573,0
C
emulovpmus
³mm63,0³mm63,0³mm0³mm63,0³mm0³mm91,0³mm0³mm91,0
*)rab(erusserP4±4124±4126±0424±4126±0424±0024±9524±002
Leakage time (waste)
Pressure must drop from 150 to 100 bars in a span of not less than 8 seconds and not more than 30.
Seat seal
Nozzle tip wet.
Pressure must be kept 20 bars below the opening pressure for 10 seconds.
After this time, dampness on the nozzle tip is acceptable, and may be identified by touching with a dry finger. Only a drop of
dampness is acceptable and not a large thick patch which would indicate a leak.
* These values apply to new injector and allow for loosening of up to10% after breaking-in period