2000-T-INS-LAB-RevC14 13
TROUBLESHOOTING
If the device does not function properly:
1. Ensure the battery is properly installed or replace the battery. Be sure
to observe proper polarity markings when replacing the battery. If the
yellow indicator light on the front of the unit no longer illuminates when
it is turned on, replace the battery and check again.
2. If the indicator light is flashing, the intensity has been adjusted, and no
stimulation is felt, check to ensure the lead wires are properly connected
and the electrodes are properly applied to the skin. If the unit appears to
be functioning and no stimulation is felt, the lead wires or electrodes may
need to be replaced.
3. If the battery appears to be charged and the unit is not functioning, turn
bothIntensityControlKnobstotheOFFposition(counterclockwise).
Then gradually turn the Intensity Control Knob (clockwise) until
stimulation is felt. If device still is not working, turn the unit off and
contact your authorized GF Health Products, Inc. distributor.
If there is any other problem, please contact an authorized GF Health
Products, Inc. distributor. Do not try to repair a defective device.